404Bedienhandbuch 1326.1578.03 ─ 10
METer<m>:CONFigure:VALue?
Returns the actual measurement type (short form) and the range.
Suffix:
<m>
.
1..4
Selects the voltmeter.
Rückgabewerte:
<Configuration> String parameter
Beispiel:
MET:CONF:VAL?
<-- "VOLT:DC 100"
The first voltmeter is set to DC voltage measurement and range
100 V.
Verwendung: Nur Abfrage
METer<m>:TRIGger:MODE <TriggerMode>
Defines how long the instrument measures. To start the measurement again, use RUN
or the appropriate METer<m>:MEASure:<function> command.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Parameter:
<TriggerMode> AUTO | SINGle
AUTO
The instrument performs continuous measurements.
SINGle
The instrument performs a single measurement.
*RST: AUTO
METer<m>:ABORt
Stops the running measurement.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Verwendung: Ereignis
Voltmeter (R&S RTH1004)