Fernsteuerbefehle
R&S
®
Scope Rider RTH
400Bedienhandbuch 1326.1578.03 ─ 10
Suffix:
<m>
.
1..4
Selects the voltmeter.
Parameter:
<Range> <numeric value> | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 1 A|10 A|100 A|
1000 A.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Bereich: 1 bis 1000
*RST: 1
Std-einheit: A
METer<m>:SENSe:VOLTage:AC:RANGe:UPPer
<Range>
METer<m>:SENSe:VOLTage:DC:RANGe:UPPer <Range>
Sets a fixed voltage range.
Suffix:
<m>
.
1..4
Selects the voltmeter.
Parameter:
<Range> <numeric value> | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 1 V|10 V|100 V|
1000 V.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Bereich: 1 bis 1000
*RST: 1
Std-einheit: V
15.8.2.4 AC+DC Strom- und Spannungsmessungen
METer<m>:SENSe:CURRent:AC:COUPling..................................................................... 400
METer<m>:SENSe:VOLTage:AC:COUPling..................................................................... 400
METer<m>:SENSe:CURRent:AC:COUPling <Coupling>
METer<m>:SENSe:VOLTage:AC:COUPling <Coupling>
Enables AC+DC measurements.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Voltmeter (R&S RTH1004)