Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
399Bedienhandbuch 1326.1578.03 ─ 10
Parameter:
<Range> <numeric value> | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 1 A|10 A|100 A|
1000 A.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Bereich: 1 bis 1000
*RST: 1
Std-einheit: A
Verwendung: Nur Einstellung
METer<m>:CONFigure:VOLTage:AC
METer<m>:CONFigure:VOLTage:DC
Sets the internal parameters and configures the range for voltage measurements.
Suffix:
<m>
.
1..4
Selects the voltmeter.
Parameter:
<Range> <numeric value> | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 1 V|10 V|100 V|
1000 V.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Bereich: 1 bis 1000
*RST: 1
Std-einheit: V
Verwendung: Nur Einstellung
15.8.2.3 METer<m>:SENSe:<Funktion>:RANGe:UPPER-Befehle
Mit METer<m>:SENSe:<function>:RANGe:UPPER-Befehlen wird der Messbereich
eingestellt. Sie können diese Befehle zusätzlich zu METer<m>:CONFigure-Befehlen
verwenden, wenn Sie nur den Bereich ändern möchten.
METer<m>:SENSe:CURRent:AC:RANGe:UPPer..............................................................399
METer<m>:SENSe:CURRent:DC:RANGe:UPPer..............................................................399
METer<m>:SENSe:VOLTage:AC:RANGe:UPPer..............................................................400
METer<m>:SENSe:VOLTage:DC:RANGe:UPPer..............................................................400
METer<m>:SENSe:CURRent:AC:RANGe:UPPer <Range>
METer<m>:SENSe:CURRent:DC:RANGe:UPPer <Range>
Sets a fixed current range.
Voltmeter (R&S RTH1004)