Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
374Bedienhandbuch 1326.1578.03 ─ 10
HARMonic:EXPort:NAME <Name>
Defines the file name, file format and path to save the harmonic measurement results.
Parameter:
<Name> String with path and file name with extension .csv.
HARMonic:EXPort:SAVE
Saves the results of the harmonic measurement to the file specified using HARMonic:
EXPort:NAME.
Beispiel:
HARMonic:EXPort:NAME 'HarmonicResults'
HARMonic:EXPort:SAVE
HARMonic:EXPort:NAME?
//Result: /media/SD/Rohde-Schwarz/RTH/Export/HarmonicResults.csv
Verwendung: Ereignis
HARMonic:RESult<m>:CLIPping?
Indicates whether clipping has occurred, that is: the amplitudes exceed the currently
defined amplitude range.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Input channel
Rückgabewerte:
<State> ON | OFF
ON
Clipping has occurred, increase the amplitude range (see
CHANnel<m>:RANGe auf Seite 303).
OFF
No clipping has occurred, the current measurement settings are
suitable.
*RST: OFF
Verwendung: Nur Abfrage
HARMonic:RESult<m>:FRQMissing?
Queries whether the fundamental frequency was detected in the input signal or not. If
the specified frequency ±10 % is not detected in the signal, the measurement is invalid.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Spektrumanalyse