Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
365Bedienhandbuch 1326.1578.03 ─ 10
Parameter:
<NumberOfMarkers> Bereich: 1 bis 15
Inkrement: 1
*RST: 3
SPECtrum:MARKer:SOURce <TraceSource>
Determines the trace used for the peak search.
Parameter:
<TraceSource> SPECtrum | MAXimum | MINimum | AVERage
*RST: SPECtrum
SPECtrum:MARKer:SETup:DISTance
<Distance>
Defines a minimum distance between two frequencies that must be exceeded in order
to detect individual peaks.
Parameter:
<Distance> Bereich: 0 bis 500E+6
Inkrement: 10
*RST: 1E+6
Std-einheit: Hz
SPECtrum:MARKer:SETup:EXCursion <Excursion>
Defines a relative threshold, the minimum level value by which the waveform must rise
or fall to be considered a peak. To avoid identifying noise peaks, enter a peak excur-
sion value that is higher than the noise levels.
See also "Auslenkung" auf Seite 130.
Parameter:
<Excursion> Bereich: 0 bis 260
Inkrement: 1
*RST: 10
Std-einheit: dB
SPECtrum:MARKer:SETup:MLEVel <Threshold>
Defines an absolute level threshold as an additional condition for the peak search.
Only peaks that exceed the threshold are detected.
Parameter:
<Threshold> Bereich: -260 bis 260
Inkrement: 1
*RST: -40
Std-einheit: dB
Spektrumanalyse