Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
362Bedienhandbuch 1326.1578.03 ─ 10
Suffix:
<m>
.
1..2
Parameter:
<State> ON | OFF
*RST: OFF
SPECtrum:CURSor<m>:SOURce <TraceSource>
Selects the trace on which the cursors are placed. Only active traces are available.
See:
SPECtrum:WAVeform:MAXimum[:ENABle] auf Seite 361
SPECtrum:WAVeform:MINimum[:ENABle] auf Seite 361
SPECtrum:WAVeform:SPECtrum[:ENABle] auf Seite 361
SPECtrum:WAVeform:AVERage[:ENABle] auf Seite 360
Suffix:
<m>
.
1..2
Parameter:
<TraceSource> SPECtrum | MAXimum | MINimum | AVERage
*RST: SPECtrum
SPECtrum:CURSor<m>:COUPling <State>
Couples the cursor lines so that the distance between the two lines remains the same
if one cursor is moved.
Suffix:
<m>
.
1..2
irrelevant
Parameter:
<State> ON | OFF
*RST: OFF
SPECtrum:CURSor<m>:SCPLing <State>
If enabled, the position of the cursor lines is adjusted if the vertical or horizontal scales
are changed. The cursor lines keep their relative position to the waveform.
If disabled, the cursor lines remain on their position on the display if the scaling is
changed.
Suffix:
<m>
.
1..2
Parameter:
<State> ON | OFF
*RST: ON
Spektrumanalyse