Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
355Bedienhandbuch 1326.1578.03 ─ 10
SPECtrum:FREQuency:STOP.........................................................................................357
SPECtrum:FREQuency:WINDow:FACTor?.......................................................................357
SPECtrum:FREQuency:WINDow:TYPE........................................................................... 357
SPECtrum:SOURce <Source>
Selects the channel for which the captured data is analyzed in FFT mode.
Parameter:
<Source> C1 | C2 | C3 | C4
*RST: C1
SPECtrum:FREQuency:BANDwidth[:RESolution][:VALue]
Queries or defines the used resolution bandwidth. The value depends on the
SPECtrum:FREQuency:SPAN:MODE and CHANnel<m>:BANDwidth parameters.
In FFT mode, this command is query only.
Parameter:
<RBW> Bereich: 1 bis 50E+9
Inkrement: 1E+6
*RST: 500E+3
Std-einheit: Hz
SPECtrum:FREQuency:CENTer <HorizCenter>
Queries or defines the used center frequency. The value depends on the SPECtrum:
FREQuency:SPAN:MODE and CHANnel<m>:BANDwidth parameters.
In FFT mode, this command is query-only.
Parameter:
<HorizCenter> Bereich: 500 bis 500E+6
Inkrement: 10
*RST: 250E+6
Std-einheit: Hz
SPECtrum:FREQuency:HORizontal:SCALe <Scaling>
Defines the scaling method for the frequency (x-)axis of the spectrum display.
Parameter:
<Scaling> LINear | LOGarithmic
*RST: LINear
SPECtrum:FREQuency:MAGNitude:SCALe <MagnitudeUnit>
Queries or defines the scale unit for the amplitude range (y-axis) in the spectrum dis-
play.
Spektrumanalyse