Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
334Bedienhandbuch 1326.1578.03 ─ 10
Parameter:
<MinDuration> Bereich: 800E-12 bis 10000
Inkrement: 100E-9
*RST: 5E-9
Std-einheit: s
TRIGger:INTerval:MAXWidth <MaxDuration>
Sets the upper time limit of the interval for comparisons WITHin and OUTSide.
Parameter:
<MaxDuration> Bereich: 800E-12 bis 10000
Inkrement: 100E-9
*RST: 5E-9
Std-einheit: s
15.3.5.15 Fenstertrigger (R&S
RTH-K19)
TRIGger:LEVel<m>:WINDow:LOWer...............................................................................334
TRIGger:LEVel<m>:WINDow:UPPer................................................................................334
TRIGger:WINDow:TIME..................................................................................................334
TRIGger:WINDow:RANGe.............................................................................................. 335
TRIGger:WINDow:WIDTh............................................................................................... 335
TRIGger:WINDow:DELTa............................................................................................... 336
TRIGger:WINDow:MINWidth........................................................................................... 336
TRIGger:WINDow:MAXWidth..........................................................................................336
TRIGger:LEVel<m>:WINDow:LOWer <Level>
TRIGger:LEVel<m>:WINDow:UPPer <Level>
Set the upper and lower voltage thresholds for the window trigger, respectively. The
trigger levels are the vertical window limits.
Suffix:
<m>
.
1..4
Indicates the trigger source: analog channel 1 to 4
Parameter:
<Level> Bereich: -10 bis 10
Inkrement: 1E-3
*RST: 0
Std-einheit: V
TRIGger:WINDow:TIME <Condition>
Selects how the time limit of the window is defined.