Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
330Bedienhandbuch 1326.1578.03 ─ 10
Parameter:
<TrigCondition> VIOLation | OK
VIOLation
Triggers on a violation of the setup or hold time
OK
Triggers if setup and hold time keep the limits.
*RST: VIOLation
TRIGger:DATatoclock:HTIMe <HoldTime>
Sets the minimum time after the clock edge while the data signal must stay steady.
The hold time can be negative. In this case, the setup time has to be positive. The
setup time is defined by TRIGger:DATatoclock:STIMe.
Parameter:
<HoldTime> Bereich: -124E-9 bis 124E-9
Inkrement: 1E-9
*RST: 1E-9
Std-einheit: s
TRIGger:DATatoclock:STIMe <SetupTime>
Sets the minimum time before the clock edge while the data signal must stay steady.
The setup time can be negative. In this case, the hold time has to be positive. The
setup time is defined by TRIGger:DATatoclock:HTIMe.
Parameter:
<SetupTime> Bereich: -124E-9 bis 124E-9
Inkrement: 1E-9
*RST: 1E-9
Std-einheit: s
15.3.5.12 Trigger für serielle Muster (R&S RTH-K19)
TRIGger:SPATtern:DSOurce[:VALue].............................................................................. 330
TRIGger:SPATtern:CSOurce[:VALue].............................................................................. 331
TRIGger:SPATtern:CSOurce:EDGE.................................................................................331
TRIGger:SPATtern:CSOurce:FIRStedge.......................................................................... 331
TRIGger:SPATtern:ORDer..............................................................................................331
TRIGger:SPATtern:PATTern...........................................................................................331
TRIGger:SPATtern:DSOurce[:VALue] <Source>
Selects the input channel of the data signal.
Parameter:
<Source> C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
*RST: C1
Grundeinstellungen