Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
327Bedienhandbuch 1326.1578.03 ─ 10
TRIGger:RUNT:MAXWidth <MaxDuration>
Sets the upper time limit of the runt for comparisons WITHin and OUTSide.
Parameter:
<MaxDuration> Bereich: 800E-12 bis 10000
Inkrement: 100E-9
*RST: 5E-9
Std-einheit: s
15.3.5.10 Anstiegszeitentrigger (R&S RTH-K19)
Siehe auch: Kapitel 3.5.10, "Anstiegszeitentrigger (R&S RTH-K19)", auf Seite 69.
TRIGger:LEVel<m>:SLEW:LOWer...................................................................................327
TRIGger:LEVel<m>:SLEW:UPPer................................................................................... 327
TRIGger:SLEW:SLOPe.................................................................................................. 327
TRIGger:SLEW:RANGe..................................................................................................327
TRIGger:SLEW:RATE.................................................................................................... 328
TRIGger:SLEW:DELTa...................................................................................................328
TRIGger:SLEW:MINWidth...............................................................................................328
TRIGger:SLEW:MAXWidth..............................................................................................329
TRIGger:LEVel<m>:SLEW:LOWer <Level>
TRIGger:LEVel<m>:SLEW:UPPer <Level>
Set the upper and lower voltage thresholds, respectively. The time measurement starts
when the signal crosses the first trigger level - the upper or lower level depending on
the selected slope - and stops when the signal crosses the second level.
Suffix:
<m>
.
1..4
Indicates the trigger source: analog channel 1 to 4
Parameter:
<Level> Bereich: -10 bis 10
Inkrement: 1E-3
*RST: 0
Std-einheit: V
TRIGger:SLEW:SLOPe <Slope>
Sets the edge direction for the trigger.
Parameter:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
TRIGger:SLEW:RANGe <Condition>
Defines the time limits of the slew rate.
Grundeinstellungen