Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
325Bedienhandbuch 1326.1578.03 ─ 10
TRIGger:LEVel<m>:RUNT:LOWer...................................................................................325
TRIGger:LEVel<m>:RUNT:UPPer................................................................................... 325
TRIGger:RUNT:POLarity................................................................................................ 325
TRIGger:RUNT:RANGe..................................................................................................325
TRIGger:RUNT:WIDTh...................................................................................................326
TRIGger:RUNT:DELTa...................................................................................................326
TRIGger:RUNT:MINWidth...............................................................................................326
TRIGger:RUNT:MAXWidth..............................................................................................327
TRIGger:LEVel<m>:RUNT:LOWer <LowerLevel>
TRIGger:LEVel<m>:RUNT:UPPer <UpperLevel>
Set the upper and lower levels that limit the runt.
Suffix:
<m>
.
1..4
Indicates the trigger source: analog channel 1 to 4
Parameter:
<LowerLevel>
<UpperLevel>
Bereich: -10 bis 10
Inkrement: 1E-3
*RST: 0
Std-einheit: V
TRIGger:RUNT:POLarity <Polarity>
Gibt die Impulspolarität an, also die Richtung der ersten Impulsflanke. Sie können trig-
gern auf:
Positive Impulse. Die Breite wird von der steigenden zur fallenden Flanke definiert.
Negative Impulse. Die Breite wird von der fallenden zur steigenden Flanke defi-
niert.
Sowohl positive als auch negative Impulse
Parameter:
<Polarity> POSitive | NEGative | EITHer
*RST: POSitive
TRIGger:RUNT:RANGe <Condition>
Defines an additional time limit of the runt pulse.
Grundeinstellungen