Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
324Bedienhandbuch 1326.1578.03 ─ 10
TRIGger:STATe:CHANnel<m> <State>
Sets the state of each input channel. The channel is specified by the channel suffix:
The logical combination of the channel states is defined by TRIGger:STATe:
COMBination.
Suffix:
<m>
.
1..4: analog channel 1 to 4
8..15: digital channels D0 to D7
Parameter:
<State> ONE | ZERO | DONTcare
ONE
The signal value is above the defined threshold.
ZERO
The signal value is below the defined threshold.
DONTcare
The signal state does not matter.
*RST: DONTcare
TRIGger:STATe:COMBination <ChCombination>
Sets the logical combination for all active channels. The required state of each channel
is defined by TRIGger:PATTern:STATe[:CHANnel<m>].
Parameter:
<ChCombination> AND | OR
*RST: AND
TRIGger:STATe:CSOurce[:VALue] <Source>
Sets the input channel of the clock signal.
Parameter:
<Source> C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
*RST: C1
TRIGger:STATe:CSOurce:EDGE <Slope>
Sets the edge of the clock at which the instrument checks the signal states.
Parameter:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
15.3.5.9 Runt-Trigger (R&S RTH-K19)
Siehe auch: Kapitel 3.5.9, "Runt-Trigger (R&S RTH-K19)", auf Seite 68.
Grundeinstellungen