Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
319Bedienhandbuch 1326.1578.03 ─ 10
Parameter:
<Mode> ALL | ODD | EVEN | ALINe | LINE
ALL
All fields: Triggers on the first video line of the frame (progres-
sive scanning) or field (interlaced scanning), for example, to find
amplitude differences between the fields.
ODD | EVEN
Odd fields / even fields: Triggers on the first video line of the odd
or even field. These modes are available for interlaced scanning
(PAL, PAL-M, SECAM, NTSC, 1080i) and progressive segmen-
ted frame scanning (1080p/24sF). They can be used, for exam-
ple, to analyze the components of a video signal.
ALINe
All lines: Triggers on the line start of all video lines, for example,
to find maximum video levels.
LINE
Triggers on a specified line. Specify the line number using
TRIGger:TV:LINE.
*RST:
ALL
TRIGger:TV:LINE <LineNumber>
Sets the number of the line to be triggered on if TRIGger:TV:MODE is LINE.
Usually the lines of the frame are counted, beginning from the frame start.
For NTSC signals, the lines are counted per field, not per frame. Therefore, you have
to set the odd or even field using TRIGger:TV:LFIeld, and the line number in the
field.
Parameter:
<LineNumber> Bereich: 1 bis 3000
Inkrement: 1
*RST: 1
TRIGger:TV:LFIeld <LineField>
Line field (odd or even) used as reference for counting the video lines. Used by the
NTSC standard.
Parameter:
<LineField> FIELD1 | FIELD2
*RST: FIELD1
TRIGger:TV:CUSTom:STYPe <SyncPulseType>
Sets the type of the sync pulse, either bi-level sync pulse (usually used in SDTV sig-
nals), or tri-level sync pulse (used in HDTV signals).
Grundeinstellungen