Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
313Bedienhandbuch 1326.1578.03 ─ 10
Parameter:
<Mode> OFF | TIME | RANDom | EVENts
OFF
No holdoff
TIME
Defines the holdoff as a time period. The next trigger occurs only
after a time has passed, which is defined with TRIGger:
HOLDoff:TIME.
RANDom
Defines the holdoff as a random time limited by TRIGger:
HOLDoff:MIN and TRIGger:HOLDoff:MAX. For each acquisi-
tion, the instrument selects a new random holdoff time from the
specified range.
EVENts
Defines the holdoff as a number of trigger events, which is defi-
ned with TRIGger:HOLDoff:EVENts.
*RST:
OFF
TRIGger:HOLDoff:TIME <Time>
Sets the time that has to pass at least until the next trigger occurs. The command
takes effect if TRIGger:MODE is set to TIME.
Parameter:
<Time> Bereich: 8E-9 bis 10
Inkrement: 200E-6
*RST: 1E-3
Std-einheit: s
TRIGger:HOLDoff:EVENts <Events>
Sets the number of triggers to be skipped until the next trigger occurs. The command
takes effect if TRIGger:MODE is set to EVENts.
Parameter:
<Events> Bereich: 1 bis 1000000000
Inkrement: 10
*RST: 1
TRIGger:HOLDoff:MIN <RandomMinTime>
TRIGger:HOLDoff:MAX <RandomMaxTime>
Set the time limits for random holdoff time. For each acquisition, the instrument selects
a new random holdoff time from the specified range.
Grundeinstellungen