Instructions
Table Of Contents
- Cover
- Contents
- 1 For Your Safety
- 2 Getting Started
- 3 Waveform Setup
- 3.1 Connecting Probes
- 3.2 Vertical Setup
- 3.3 Horizontal Setup
- 3.4 Acquisition Control
- 3.5 Roll Mode
- 3.6 Trigger
- 3.6.1 General Trigger Settings
- 3.6.2 Edge Trigger
- 3.6.3 Glitch Trigger
- 3.6.4 Width Trigger
- 3.6.5 Video/TV Trigger
- 3.6.6 External Trigger (R&S RTH1002)
- 3.6.7 Pattern Trigger (R&S RTH-K19)
- 3.6.8 State Trigger (R&S RTH-K19)
- 3.6.9 Runt Trigger (R&S RTH-K19)
- 3.6.10 Slew Rate Trigger (R&S RTH-K19)
- 3.6.11 Data2Clock Trigger (R&S RTH-K19)
- 3.6.12 Serial Pattern Trigger (R&S RTH-K19)
- 3.6.13 Timeout Trigger (R&S RTH-K19)
- 3.6.14 Interval Trigger (R&S RTH-K19)
- 3.6.15 Window Trigger (R&S RTH-K19)
- 3.6.16 Protocol Trigger (R&S RTH-K1, -K2, -K3, -K9 and -K10)
- 4 Waveform Analysis
- 5 Mask Testing
- 6 Spectrum Analysis
- 6.1 FFT Mode
- 6.2 Spectrum Mode (Option R&S RTH-K18)
- 6.3 Harmonics Measurement (Option R&S RTH-K34)
- 7 Multimeter Measurements
- 8 Data Logging
- 9 Protocol Analysis
- 10 Logic Analyzer (R&S RTH-B1 MSO)
- 11 Frequency Counter (R&S RTH-K33)
- 12 Documenting Results
- 13 General Instrument Setup
- 14 Network Connections
- 15 Remote Control Commands
- 15.1 Conventions used in Command Description
- 15.2 Mode
- 15.3 Waveform Setup
- 15.3.1 Automatic Setup
- 15.3.2 Vertical Setup
- 15.3.3 Horizontal Setup
- 15.3.4 Acquisition Control
- 15.3.5 Trigger
- 15.3.5.1 General Trigger Settings
- 15.3.5.2 Edge Trigger
- 15.3.5.3 Glitch Trigger
- 15.3.5.4 Width Trigger
- 15.3.5.5 Video/TV Trigger
- 15.3.5.6 External Trigger (R&S RTH1002)
- 15.3.5.7 Pattern Trigger (R&S RTH-K19)
- 15.3.5.8 State Trigger (R&S RTH-K19)
- 15.3.5.9 Runt Trigger (R&S RTH-K19)
- 15.3.5.10 Slew Rate Trigger (R&S RTH-K19)
- 15.3.5.11 Data2Clock Trigger (R&S RTH-K19)
- 15.3.5.12 Serial Pattern Trigger (R&S RTH-K19)
- 15.3.5.13 Timeout Trigger (R&S RTH-K19)
- 15.3.5.14 Interval Trigger (R&S RTH-K19)
- 15.3.5.15 Window Trigger (R&S RTH-K19)
- 15.4 Waveform Analysis
- 15.5 Mask Testing
- 15.6 Spectrum Analysis
- 15.7 Digital Multimeter (R&S RTH1002)
- 15.8 Voltmeter (R&S RTH1004)
- 15.9 Counter Mode (R&S RTH-K33)
- 15.10 Data Logging
- 15.11 Protocol Analysis
- 15.12 Logic Analyzer (R&S RTH-B1 MSO)
- 15.13 Documenting Results
- 15.14 General Instrument Setup
- 15.15 WLAN Connection (Option R&S RTH-K200/200US)
- 15.16 User Scripting (R&S RTH-K38)
- Annex
- List of Commands
Waveform Analysis
R&S
®
Scope Rider RTH
95User Manual 1326.1578.02 ─ 15
Remote command:
REFCurve:STATe on page 348
CALCulate:MATH:STATe on page 346
Vertical Position
Sets the vertical position of the reference waveform.
You can also tap the waveform label "R" to set the focus to the reference waveform,
and use the [RANGE] and [POS] keys to adjust the display.
Remote command:
REFCurve:POSition on page 348
Save/ Load
Saves or loads a reference waveform. The default directory is C:/Users/<user>/
Rohde-Schwarz/RTH/ReferenceCurves.
Select the "File Type" (format BIN, XML, or CSV) and enter the "File Name". See also
Chapter 4.5.2, "Waveform Files", on page 95.
Remote command:
REFCurve:NAME on page 348
REFCurve:SAVE on page 348
REFCurve:OPEN on page 348
REFCurve:DELete on page 348
4.5.2 Waveform Files
Reference waveforms can be stored in XML, CSV, or BIN format.
If you want to reload reference waveforms on the instrument, save them in BIN or CSV
format. XML files cannot be reloaded.
Waveform data is saved in two files. One file contains the waveform data values and is
indicated by *Wfm.* in the file name. The second file contains the header data, for
example, time scale, vertical scale, vertical position, acquisition mode, and more.
Header data is required to reload the waveform from data, or to analyze the data val-
ues of the data file.
4.5.2.1 Waveform Header Files
The header files of XML and BIN waveform files are written in XML format. The header
files of CSV waveform files are written in CSV format. You can open the header files
and use their information for data analysis.
CSV header files only contain the property names and values, one property per row.
VerticalScale:0.05:
HorizontalScale:5e-08:
Reference Waveforms