Instructions

Table Of Contents
Waveform Analysis
R&S
®
Scope Rider RTH
91User Manual 1326.1578.02 ─ 15
Show Position
Shows the position values of the vertical cursor lines t1 and t2 in the measurement
results. The setting is only available for the vertical cursor type.
Meas Type 1 / Meas Type 2
Set the automatic measurements to be performed on the source waveform between
the cursor lines. The setting is only available for the "Measure" cursor type.
All automatic measurements that need only one source are available. Delay, phase,
and power measurements are not provided for cursor measurements, they require two
sources.
For a description of the measurement types, see Chapter 4.2.3, "Measurement Types",
on page 81.
Remote command:
CURSor:MEASurement<m>:TYPE on page 343
Track Scaling
If enabled, the position of the cursor lines is adjusted if the vertical or horizontal scales
are changed. The cursor lines keep their relative position to the waveform.
If disabled, the cursor lines remain on their position on the display if the scaling is
changed.
Remote command:
CURSor:SCPLing on page 343
Coupling
Couples the cursor lines so that the distance between the two lines remains the same
if one cursor is moved.
Remote command:
CURSor:COUPling on page 343
Set to Screen
Sets the cursors to a default position on the screen. This is helpful if the cursors have
disappeared from the display or need to be moved for a larger distance.
Remote command:
CURSor:SCReen on page 343
4.4 Mathematics
A math waveform is a calculated waveform. You can calculate data out of one or two
different sources using several predefined operations.
To configure the math waveform, press and hold the [MATH] key until the menu
opens.
Mathematics