Instructions

Table Of Contents
Waveform Analysis
R&S
®
Scope Rider RTH
90User Manual 1326.1578.02 ─ 15
State
Enables or disables the cursor measurement.
Remote command:
CURSor:STATe on page 342
Type
Defines the type of the cursor measurement.
For details, see Chapter 4.3.2, "Cursor Types and Results", on page 88.
"Vertical"
Displays two vertical cursor lines and measures their timing parame-
ters.
"Horizontal"
Displays two horizontal cursor lines and measures their amplitude
parameters.
"Track"
Displays two vertical cursor lines and couples them to the source
waveform. The amplitude characteristcs and the time difference of the
crossing points are measured.
"Measure"
Displays two vertical cursor lines that define a gate for two simultane-
ous automatic measurements.
Remote command:
CURSor:FUNCtion on page 342
Source
Defines the source on which the cursor measurement is performed. The source can be
any active analog or digital input signal, math waveform, or bus (requires option).
The source setting is not available for the cursor type "Vertical", and for measurements
in the XY-diagram.
Remote command:
CURSor:SOURce on page 342