Instructions

Table Of Contents
Waveform Analysis
R&S
®
Scope Rider RTH
88User Manual 1326.1578.02 ─ 15
4.3.1 Performing Cursor Measurements
To start and stop the last configured measurement
Press the [Cursor] key.
To configure the cursor measurement in the Cursor menu
1. Press and hold the [Cursor] key until the "Cursor" menu opens.
2. Select the "Type" of the cursor.
3. For horizontal, track and measure types, select the "Source" channel that you want
to measure.
4. Scroll down the menu and adjust the additional settings, which are required for the
selected cursor type.
4.3.2 Cursor Types and Results
The results of cursor measurements are displayed at the top of the display. 4 cursor
types are available.
Vertical cursors
For vertical cursors, two results are displayed by default: the absolute value of the time
difference between the cursor lines Δt and its inverse value 1/Δt. Optionally, the posi-
tions of the cursor lines t1 and t2 are also measured. The results are time values and
do not depend on any waveform, thus no source is required.
CURSor:TDELta? on page 344
CURSor:ITDelta? on page 344
CURSor:X1Position on page 344
CURSor:X2Position on page 344
Horizontal cursors
For horizontal cursors, the vertical values of the cursor positions y1 and y2 are mea-
sured. These are usually voltage or current values. The absolute value of the differ-
ence between the positions Δy is also displayed.
CURSor:Y1Position on page 345
CURSor:Y2Position on page 345
CURSor:DELTa? on page 344
Cursor Measurements