Instructions

Table Of Contents
Waveform Analysis
R&S
®
Scope Rider RTH
87User Manual 1326.1578.02 ─ 15
Remote command:
MEASurement<m>:ENABle on page 339
Type
Selects the measurement type. For a detailed description, see Chapter 4.2.3, "Mea-
surement Types", on page 81.
Remote command:
MEASurement<m>:TYPE on page 340
Source / Source 2
Defines the waveform to be measured. For delay, phase, and power measurements, 2
sources are required.
The sources can be any active input signal, math or reference waveform. Available
source waveforms depend on the measurement type, see Chapter 4.2.3, "Measure-
ment Types", on page 81.
Remote command:
MEASurement<m>:SOURce on page 340
All off
Disables all active measurements.
Remote command:
MEASurement<m>:AOFF on page 341
Slope
Sets the slope for the delay measurement type.
"Positive"
Delay between the first rising edge of each source waveform.
"Negative"
Delay between the first falling edge of each source waveform.
"Either"
Delay between the first edge of each source waveform, no matter if it
is rising or falling.
Remote command:
MEASurement<m>:DELay:SLOPe on page 341
4.3 Cursor Measurements
The cursor measurement determines the results at the current cursor positions, or per-
forms gated automatic measurements between the cursor lines. The cursors can be
positioned manually at fixed positions, or they can follow the waveform.
You can perform cursor measurements on analog input signals, math waveform, XY-
diagram, as well as on logic channels (requires option R&S RTH-B1).
Cursor measurements on spectrum displays are also available, if the option R&S RTH-
K18 is installed (see Chapter 6.2.7, "Cursor Measurements on Spectrums",
on page 128).
Cursor Measurements