R&S
®
Scope Rider RTH
80User Manual 1326.1578.02 ─ 15
Remote command:
ZOOM:POSition on page 339
4.2 Automatic Measurements
You can perform up to four different measurements simultaneously.
4.2.1 Performing Automatic Measurements
To start and stop the last configured measurements
Press the [Meas] key.
To configure automatic measurements in the Meas menu
1. Press and hold the [Meas] key until the "Meas" menu opens.
2. Select the number of the measurement that you want to configure.
3. If the measurement is disabled, enable "State".
4. Select the "Type".
The selection list shows all available measurement types.
5. Select the "Source".
The selection list shows all active sources that are allowed for the selected mea-
surement type.
6. Some measurement types require additional settings. Scroll down the menu and
adjust the additional settings if necessary.
4.2.2 Measurement Results
The measurement results are shown on the left-top side of the screen.
Figure 4-3: Results of 4 active measurements
If a result cannot be determined, "---" is displayed. Adjust the horizontal and vertical
settings if the instrument cannot measure.
If the measurement result is outside the measurement range and clipping occurs, the
results are marked with < (underflow) or > (overflow). Adjust the vertical scale to get
valid results.
Automatic Measurements