Instructions

Table Of Contents
Waveform Setup
R&S
®
Scope Rider RTH
70User Manual 1326.1578.02 ─ 15
Description of settings
Figure 3-16: Data2Clock trigger
Clock Source
Selects the input channel of the clock signal.
Remote command:
TRIGger:DATatoclock:CSOurce[:VALue] on page 331
Clock Slope
Sets the edge of the clock signal: rising, falling, or both edges. The time reference
point for the setup and hold time is the crossing point of the clock edge and the trigger
level.
Remote command:
TRIGger:DATatoclock:CSOurce:EDGE on page 331
Data Source
Selects the input channel of the data signal.
Remote command:
TRIGger:DATatoclock:DSOurce[:VALue] on page 331
Trigger on Setup & Hold
Selects how a violation of the setup and hold time is handled.
"Violation"
Triggers on a violation of the setup or hold time
"OK"
Triggers if setup and hold time keep the limits.
Remote command:
TRIGger:DATatoclock:CONDition on page 331
Trigger