Instructions
Table Of Contents
- Cover
- Contents
- 1 For Your Safety
- 2 Getting Started
- 3 Waveform Setup
- 3.1 Connecting Probes
- 3.2 Vertical Setup
- 3.3 Horizontal Setup
- 3.4 Acquisition Control
- 3.5 Roll Mode
- 3.6 Trigger
- 3.6.1 General Trigger Settings
- 3.6.2 Edge Trigger
- 3.6.3 Glitch Trigger
- 3.6.4 Width Trigger
- 3.6.5 Video/TV Trigger
- 3.6.6 External Trigger (R&S RTH1002)
- 3.6.7 Pattern Trigger (R&S RTH-K19)
- 3.6.8 State Trigger (R&S RTH-K19)
- 3.6.9 Runt Trigger (R&S RTH-K19)
- 3.6.10 Slew Rate Trigger (R&S RTH-K19)
- 3.6.11 Data2Clock Trigger (R&S RTH-K19)
- 3.6.12 Serial Pattern Trigger (R&S RTH-K19)
- 3.6.13 Timeout Trigger (R&S RTH-K19)
- 3.6.14 Interval Trigger (R&S RTH-K19)
- 3.6.15 Window Trigger (R&S RTH-K19)
- 3.6.16 Protocol Trigger (R&S RTH-K1, -K2, -K3, -K9 and -K10)
- 4 Waveform Analysis
- 5 Mask Testing
- 6 Spectrum Analysis
- 6.1 FFT Mode
- 6.2 Spectrum Mode (Option R&S RTH-K18)
- 6.3 Harmonics Measurement (Option R&S RTH-K34)
- 7 Multimeter Measurements
- 8 Data Logging
- 9 Protocol Analysis
- 10 Logic Analyzer (R&S RTH-B1 MSO)
- 11 Frequency Counter (R&S RTH-K33)
- 12 Documenting Results
- 13 General Instrument Setup
- 14 Network Connections
- 15 Remote Control Commands
- 15.1 Conventions used in Command Description
- 15.2 Mode
- 15.3 Waveform Setup
- 15.3.1 Automatic Setup
- 15.3.2 Vertical Setup
- 15.3.3 Horizontal Setup
- 15.3.4 Acquisition Control
- 15.3.5 Trigger
- 15.3.5.1 General Trigger Settings
- 15.3.5.2 Edge Trigger
- 15.3.5.3 Glitch Trigger
- 15.3.5.4 Width Trigger
- 15.3.5.5 Video/TV Trigger
- 15.3.5.6 External Trigger (R&S RTH1002)
- 15.3.5.7 Pattern Trigger (R&S RTH-K19)
- 15.3.5.8 State Trigger (R&S RTH-K19)
- 15.3.5.9 Runt Trigger (R&S RTH-K19)
- 15.3.5.10 Slew Rate Trigger (R&S RTH-K19)
- 15.3.5.11 Data2Clock Trigger (R&S RTH-K19)
- 15.3.5.12 Serial Pattern Trigger (R&S RTH-K19)
- 15.3.5.13 Timeout Trigger (R&S RTH-K19)
- 15.3.5.14 Interval Trigger (R&S RTH-K19)
- 15.3.5.15 Window Trigger (R&S RTH-K19)
- 15.4 Waveform Analysis
- 15.5 Mask Testing
- 15.6 Spectrum Analysis
- 15.7 Digital Multimeter (R&S RTH1002)
- 15.8 Voltmeter (R&S RTH1004)
- 15.9 Counter Mode (R&S RTH-K33)
- 15.10 Data Logging
- 15.11 Protocol Analysis
- 15.12 Logic Analyzer (R&S RTH-B1 MSO)
- 15.13 Documenting Results
- 15.14 General Instrument Setup
- 15.15 WLAN Connection (Option R&S RTH-K200/200US)
- 15.16 User Scripting (R&S RTH-K38)
- Annex
- List of Commands
Waveform Setup
R&S
®
Scope Rider RTH
64User Manual 1326.1578.02 ─ 15
Figure 3-12: Pattern definition
The current threshold is displayed for each channel. For analog channels, the thresh-
old is set in the "Vertical" menu > "Technology". For logic channels, the threshold is set
in the "Logic" menu. To adjust the thresholds by the instrument, use "Find Level".
The switches define the state of each channel and set the logical combination:
"1"
The signal value is above the defined threshold.
"0"
The signal value is below the defined threshold.
"X"
The signal state does not matter.
"AND"
If all defined states are true, the logical result of the pattern definition
is 1 (true).
"OR"
If at least one of the defined states is true, the logical result of the pat-
tern definition is 1 (true).
Remote command:
TRIGger:PATTern:STATe[:CHANnel<m>] on page 323
TRIGger:PATTern:STATe:COMBination on page 323
Range
Adds additional time limitation to the defined pattern.
"None"
No time limit is set. If the defined pattern is true, the pattern trigger
occurs.
"Timeout"
Defines a minimum time during which the signals match the pattern
definition.
"Longer"
If the pattern is true longer than the "Pattern Width" time, the trigger
occurs.
"Shorter"
If the pattern is true for a time shorter than "Pattern Width", the trigger
occurs.
"Equal"
If the pattern is true for the time "Pattern Width"
±
Δt ("Tolerance"),
the trigger occurs.
Trigger