Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
488User Manual 1326.1578.02 ─ 15
Suffix:
<m>
.
*
Return values:
<StatusCommunication>Range: 0 to 0
Increment: 1
*RST: 0
Usage: Query only
BUS:SENT:FRAMe<m>:SDATa?
Returns the symbolic data of the frame.
Suffix:
<m>
.
*
Return values:
<SymbolicData> Comma-separated list of values. The first value is the number of
bytes, followed by the decoded data bytes.
Usage: Query only
BUS:SENT:FRAMe<m>:STARt?
Returns the start time of the selected frame.
Suffix:
<m>
.
*
Return values:
<FrameStart> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS:SENT:FRAMe<m>:STATus?
Returns the overall state of the selected frame.
Suffix:
<m>
.
*
Return values:
<FrameState> OK | SYNC | PULSe | CRC | IRFL | FORM | INSufficient
OK: The frame is valid.
SYNC: Synchronization error occurred.
PULse: Pulse error occurred.
CRC: Cyclic redundancy check failed.
IRFL: Irregular frame length error occurred.
FORM: Format error occurred.
Protocol Analysis