TRIGger:SENT:ICONdition <Relation>
Defines how the specified identifier pattern is compared with the acquired signal. The
instrument triggers if the acquired address is equal or unequal the defined pattern.
Parameters:
<Relation> UNUSed | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
*RST: EQUal
TRIGger:SENT:IDENtifier <Pattern>
Specifies the identifier pattern to be found, in binary format. Enter the pattern in MSB
first bit order.
Parameters:
<Pattern> String with max. 7 characters. Characters 0, 1, and X are
allowed. If you define a pattern shorter than the pattern length,
the missing LSB are filled with X.
TRIGger:SENT:SCONdition <Relation>
Defines how the specified status pattern is compared with the acquired signal. The
instrument triggers if the acquired address is equal or unequal the defined pattern.
Parameters:
<Relation> UNUSed | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
*RST: EQUal
TRIGger:SENT:STATus <Pattern>
Defines the status pattern as trigger condition. Enter the words in MSB first bit order.
Parameters:
<Pattern>
TRIGger:SENT:FCRCerror <FastCRCError>
Triggers on a detected CRC error in the transmission sequences of the fast channel.
The CRC length is 4 bits.
Parameters:
<FastCRCError> ON | OFF
*RST: ON
Protocol Analysis