Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
480User Manual 1326.1578.02 ─ 15
BUS:SENT:CLKPeriod <ClockPeriod>
Sets the nominal clock period (clock tick).
Parameters:
<ClockPeriod> Range: 1E-6 to 100E-6
Increment: 1E-6
*RST: 3E-6
Default unit: s
BUS:SENT:CLKTolerance
<ClockTolerance>
Sets a tolerated deviation of the clock signal.
Parameters:
<ClockTolerance> Range: 0 to 25
Increment: 1
*RST: 20
Default unit: %
BUS:SENT:CRCMethod <CRCCalculation>
Selects the calculation method for the CRC checksum.
Parameters:
<CRCCalculation> SAEJ | TLE
SAEJ: according to the standard
TLE: according to the computing method for TLE_4998X sen-
sors.
*RST: SAEJ
BUS:SENT:CRCVersion <CRCVersion>
Selects the calculaction method for the cyclic redundancy check (CRC).
Parameters:
<CRCVersion> V2010 | LEGA
LEGAcy: method used up to 2010
V2010: current method
*RST: V2010
BUS:SENT:DNIBbles <DataNibbles>
Sets the number of data nibbles for a transmission sequence.
Parameters:
<DataNibbles> Range: 1 to 6
Increment: 1
*RST: 3
Protocol Analysis