Instructions
Table Of Contents
- Cover
- Contents
- 1 For Your Safety
- 2 Getting Started
- 3 Waveform Setup
- 3.1 Connecting Probes
- 3.2 Vertical Setup
- 3.3 Horizontal Setup
- 3.4 Acquisition Control
- 3.5 Roll Mode
- 3.6 Trigger
- 3.6.1 General Trigger Settings
- 3.6.2 Edge Trigger
- 3.6.3 Glitch Trigger
- 3.6.4 Width Trigger
- 3.6.5 Video/TV Trigger
- 3.6.6 External Trigger (R&S RTH1002)
- 3.6.7 Pattern Trigger (R&S RTH-K19)
- 3.6.8 State Trigger (R&S RTH-K19)
- 3.6.9 Runt Trigger (R&S RTH-K19)
- 3.6.10 Slew Rate Trigger (R&S RTH-K19)
- 3.6.11 Data2Clock Trigger (R&S RTH-K19)
- 3.6.12 Serial Pattern Trigger (R&S RTH-K19)
- 3.6.13 Timeout Trigger (R&S RTH-K19)
- 3.6.14 Interval Trigger (R&S RTH-K19)
- 3.6.15 Window Trigger (R&S RTH-K19)
- 3.6.16 Protocol Trigger (R&S RTH-K1, -K2, -K3, -K9 and -K10)
- 4 Waveform Analysis
- 5 Mask Testing
- 6 Spectrum Analysis
- 6.1 FFT Mode
- 6.2 Spectrum Mode (Option R&S RTH-K18)
- 6.3 Harmonics Measurement (Option R&S RTH-K34)
- 7 Multimeter Measurements
- 8 Data Logging
- 9 Protocol Analysis
- 10 Logic Analyzer (R&S RTH-B1 MSO)
- 11 Frequency Counter (R&S RTH-K33)
- 12 Documenting Results
- 13 General Instrument Setup
- 14 Network Connections
- 15 Remote Control Commands
- 15.1 Conventions used in Command Description
- 15.2 Mode
- 15.3 Waveform Setup
- 15.3.1 Automatic Setup
- 15.3.2 Vertical Setup
- 15.3.3 Horizontal Setup
- 15.3.4 Acquisition Control
- 15.3.5 Trigger
- 15.3.5.1 General Trigger Settings
- 15.3.5.2 Edge Trigger
- 15.3.5.3 Glitch Trigger
- 15.3.5.4 Width Trigger
- 15.3.5.5 Video/TV Trigger
- 15.3.5.6 External Trigger (R&S RTH1002)
- 15.3.5.7 Pattern Trigger (R&S RTH-K19)
- 15.3.5.8 State Trigger (R&S RTH-K19)
- 15.3.5.9 Runt Trigger (R&S RTH-K19)
- 15.3.5.10 Slew Rate Trigger (R&S RTH-K19)
- 15.3.5.11 Data2Clock Trigger (R&S RTH-K19)
- 15.3.5.12 Serial Pattern Trigger (R&S RTH-K19)
- 15.3.5.13 Timeout Trigger (R&S RTH-K19)
- 15.3.5.14 Interval Trigger (R&S RTH-K19)
- 15.3.5.15 Window Trigger (R&S RTH-K19)
- 15.4 Waveform Analysis
- 15.5 Mask Testing
- 15.6 Spectrum Analysis
- 15.7 Digital Multimeter (R&S RTH1002)
- 15.8 Voltmeter (R&S RTH1004)
- 15.9 Counter Mode (R&S RTH-K33)
- 15.10 Data Logging
- 15.11 Protocol Analysis
- 15.12 Logic Analyzer (R&S RTH-B1 MSO)
- 15.13 Documenting Results
- 15.14 General Instrument Setup
- 15.15 WLAN Connection (Option R&S RTH-K200/200US)
- 15.16 User Scripting (R&S RTH-K38)
- Annex
- List of Commands
Remote Control Commands
R&S
®
Scope Rider RTH
479User Manual 1326.1578.02 ─ 15
BUS:SENT:CRCMethod..................................................................................................480
BUS:SENT:CRCVersion..................................................................................................480
BUS:SENT:DNIBbles......................................................................................................480
BUS:SENT:PPFLength................................................................................................... 481
BUS:SENT:PPULse........................................................................................................481
BUS:SENT:SFORmat..................................................................................................... 481
BUS:SENT:RDSL...........................................................................................................481
BUS:SENT:DATA:SOURce <Channel>
Sets the source of the data line. All channel waveforms can be used.
Parameters:
<Channel> C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
C3 and C4: only R&S RTH1004
Digital channels are available if option R&S RTH-B1 is installed.
*RST: C1
Usage: Asynchronous command
BUS:SENT:POLarity <Polarity>
Sets the idle state.
Parameters:
<Polarity> IDLLow | IDLHigh
*RST: IDLHigh
BUS:SENT:TECHnology <ThresholdType>
Sets the threshold values as defined for various signal technologies.
Parameters:
<ThresholdType> CMOS | USER
*RST: USER
BUS:SENT:DATA:THReshold <ThresholdValue>
Sets the threshold value for digitization of signals manually. If the signal value is higher
than the threshold, the signal state is high. Otherwise, the signal state is considered
low.
This value is only considered for BUS:SENT:TECHnology on page 479 USER.
Parameters:
<ThresholdValue> Range: -400 to 400
Increment: 1E-3
*RST: 1.4
Default unit: V
Protocol Analysis