Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
477User Manual 1326.1578.02 ─ 15
Return values:
<IdentifierValue> Range: 0 to 63
*RST: 0
Usage: Query only
BUS:LIN:FRAMe<m>:STARt?
BUS:LIN:FRAMe<m>:STOP?
Returns the start time and stop time of the selected frame, respectively.
Suffix:
<m>
.
*
Return values:
<FrameStart>
<FrameStop>
Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS:LIN:FRAMe<m>:STATus?
Returns the overall state of the selected frame.
Suffix:
<m>
.
*
Return values:
<FrameState> OK | WAKeup | SYERror | PRERror | CHCKsum | CPERror |
INSufficient
CHCKsum: checksum error
PRERror: parity error in identifier
SYERror: synchronization error
WAKeup: the frame is a wakeup frame
CPERror: parity error and checksum error
INSufficient: the frame is not completely contained in the acqui-
sition. The decoded part of the frame is valid.
*RST: OK
Usage: Query only
BUS:LIN:FRAMe<m>:SYMBol?
Returns the symbolic label of the specified frame if the label list is enabled.
Suffix:
<m>
.
*
Return values:
<Label> String with symbolic name of the identifier
Protocol Analysis