Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
470User Manual 1326.1578.02 ─ 15
Parameters:
<Channel> C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
C3 and C4: only R&S RTH1004
Digital channels are available if option R&S RTH-B1 is installed.
*RST: C1
Usage: Asynchronous command
BUS:LIN:POLarity <Polarity>
Defines the idle state of the bus. The idle state is the recessive state and corresponds
to a logical 1.
Parameters:
<Polarity> IDLLow | IDLHigh
IDLLow
The bus is idle (state = 1) when the signal is low
IDLHigh
The bus is idle (state = 1) when the signal is high
*RST: IDLHigh
BUS:LIN:BITRate <CustomBitrate>
Sets the number of transmitted bits per second.
Parameters:
<CustomBitrate> Range: 1000 to 20000
Increment: 1
*RST: 9600
Default unit: bps
BUS:LIN:STANdard <Standard>
Selects the version of the LIN standard that is used in the DUT. The setting mainly
defines the checksum version used during decoding.
Parameters:
<Standard> V1X | V2X | J2602 | AUTO
*RST: AUTO
BUS:LIN:DATA:THReshold <ThresholdValue>
Sets the threshold value for digitization of signals manually. If the signal value is higher
than the threshold, the signal state is high. Otherwise, the signal state is considered
low.
This value is only considered for BUS:LIN:TECHnology USER
Protocol Analysis