Instructions

Table Of Contents
Waveform Setup
R&S
®
Scope Rider RTH
47User Manual 1326.1578.02 ─ 15
You can set the trigger point even outside the diagram and analyze the signal some
time before or after the trigger. In this case, the trigger marker is shown on the left or
right side of the diagram.
To set the horizontal position, you can also use the [POS] keys.
Remote command:
TIMebase:HORizontal:POSition on page 309
Reference Point
Defines the time reference point in the diagram. You can set the reference point in the
middle, or to the right to see the signal before the trigger. If the reference point is on
the left, you see the signal after the trigger.
Remote command:
TIMebase:REFerence on page 310
3.4 Acquisition Control
Acquisition settings define the processing of the captured samples in the instrument.
To adjust the acquisition settings, press the [ACQUIRE] key, or select "Acquire" in
the main menu.
To start or stop acquisition, press the [RUN STOP] key.
The R&S RTH captures the input signal and converts it to digital samples. The digital
samples are processed according to the acquisition settings. The result is a waveform
record that is displayed on the screen and stored in memory.
Acquisition Control