Instructions

Table Of Contents
Waveform Setup
R&S
®
Scope Rider RTH
46User Manual 1326.1578.02 ─ 15
T
Reference point
= Left
Reference point
= Middle
Reference point
= Right
t = 0
Horiz. position from
left ref. point (< 0)
Horiz. position from
middle ref. point (> 0)
Horiz. position from right ref. point (> 0)
Signal delay
If you have set a bandwidth limit, the signal might appear delayed on the screen. The
delay time depends on the selected filter. The effect is visible if several signals with dif-
ferent limits are displayed.
See also: Chapter 3.2.2, "Effect of the Bandwidth Filter", on page 45.
Description of settings
Time Scale
Sets the time scale (timebase) of the horizontal axis for all signals, in seconds per divi-
sion. The value is shown in the top information bar.
Increase the scale to see a longer part of the waveform. Decrease the scale to see the
signal in more detail. The scale has a point that remains fixed on the screen when the
scale value is changing - the reference point.
To set the time scale, you can also use the [TIME] keys.
Note: In FFT mode, the time scale may be restricted depending on the selected fre-
quency span, (see "Frequency Span" on page 113).
Remote command:
TIMebase:SCALe on page 309
TIMebase:RANGe on page 309
Horizontal Position
Sets the horizontal position of the trigger point in relation to the reference point. The
trigger position is marked by a colored triangle at the top of the diagram.
Horizontal Setup