Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
459User Manual 1326.1578.02 ─ 15
TRIGger:CAN:CRCerror <CrcError>
A CRC error occurs when the calculated result differs from the received value in the
CRC sequence.
Parameters:
<CrcError> ON | OFF
*RST: ON
TRIGger:CAN:DATA <Pattern>
Specifies the data pattern to be found, in binary or hex format. Enter the pattern in
MSB first bit order.
Parameters:
<Pattern> String that contains the pattern in binary format. The parameter
accepts the bit value X (don't care).
TRIGger:CAN:DCONdition <Relation>
Defines how the specified data pattern is compared with the acquired signal. The
instrument triggers if the acquired address is equal or unequal the defined pattern.
Parameters:
<Relation> EQUal | NEQual
*RST: EQUal
TRIGger:CAN:FORMerror <FormError>
A form error occurs when a fixed-form bit field contains one or more illegal bits.
Parameters:
<FormError> ON | OFF
*RST: ON
TRIGger:CAN:FTYPe <FrameType>
CAN has several frame types which can be used as trigger condition.
For data and remote frames, the identifier format has to be set with TRIGger:CAN:
ITYPe.
Parameters:
<FrameType> ERRor | OVERload | DATA | REMote | DOR
ERRor
When a node recognizes an error, it cancels transmission by
sending an error frame.
The instrument triggers seven bit periods after the end of the
error flag that is marked by a dominant-recessive edge.
Protocol Analysis