452User Manual 1326.1578.02 ─ 15
TRIGger:UART:DATA <Pattern>
Defines the data pattern as trigger condition. Enter the words in MSB first bit order.
Parameters:
<Pattern> String with max. 8 bit in binary format. Characters 0, 1, and X
are allowed.
Example:
TRIGger:UART:DATA '1x11'
TRIGger:UART:DATA?
1X11
TRIGger:UART:DCONdition <Relation>
Defines how the specified data pattern is compared with the acquired signal. The
instrument triggers if the acquired data is equal or unequal to the pattern.
Parameters:
<Relation> EQUal | NEQual
*RST: EQUal
15.11.4.3 UART Decode Results
The frame suffix <m> selects the frame index for which the result is queried.
The byte suffix <n> selects the byte index for which the result is queried.
BUS:UART:WCOunt?..................................................................................................... 452
BUS:UART:WORD<m>:STATe?.......................................................................................452
BUS:UART:WORD<m>:STARt?.......................................................................................453
BUS:UART:WORD<m>:STOP?....................................................................................... 453
BUS:UART:WORD<m>:VALue?...................................................................................... 454
BUS:UART:WCOunt?
Returns the number of decoded symbols (words).
Return values:
<Count> Total number of decoded words.
Range: 0 to 100000
Increment: 1
*RST: 0
Usage: Query only
BUS:UART:WORD<m>:STATe?
Returns the status of the specified symbol (word).
Protocol Analysis