450User Manual 1326.1578.02 ─ 15
BUS:UART:BITRate <CustomBitrate>
Sets a user-defined bit rate if BUS:UART:STDBitrate is set to CUSTom.
Parameters:
<CustomBitrate> Range: 300 to 20000000
Increment: 1
*RST: 14400
Default unit: bps
BUS:UART:SSIZe
<DataBits>
Sets the number of data bits in a word (symbol).
Parameters:
<DataBits> B5 | B6 | B7 | B8 | B9
*RST: B8
BUS:UART:PARity <Parity>
Defines the optional parity bit that is used for error detection.
Parameters:
<Parity> NONE | ODD | EVEN
NONE
No parity bit is used.
ODD
The parity bit is set to "1" if the number of data bits set to "1" is
even.
EVEN
The parity bit is set to "1" if the number of data bits set to "1" is
odd.
*RST: NONE
BUS:UART:SBIT <StopBits>
Sets the number of stop bits: 1 or 1.5 or 2 stop bits are possible.
Parameters:
<StopBits> B1 | B15 | B2
*RST: B1
BUS:UART:ORDer <BitOrder>
Defines if a word starts with MSB (most significant bit) or LSB (least significant bit).
The display of the decoded signal considers this setting, results are displayed in the
specified order.
Protocol Analysis