Instructions

Table Of Contents
Waveform Setup
R&S
®
Scope Rider RTH
44User Manual 1326.1578.02 ─ 15
Remote command:
CHANnel<m>:DESKew on page 307
Technology, Value
Sets the threshold, which is used to obtain the signal state. If the signal value is higher
than the threshold, the signal state is high (1 or true for the Boolean logic). Otherwise,
if the signal value is below the threshold, the signal state is considered low (0 or false).
The threshold is used by the pattern and state trigger.
If a protocol option is installed, and the channel is used in the bus, the configured
channel threshold is also used in the bus configuration. The values are the same in
"Vertical" menu and bus configuration dialogs.
"Technology"
Select a predefined value for one of the most common technologies,
or select "User" to define an individual threshold.
"Value"
Set an individual threshold value if "Technology" is set to "User".
Remote command:
CHANnel<m>:THReshold:TECHnology on page 308
CHANnel<m>:THReshold:USER on page 308
CHANnel<m>:THReshold:THReshold? on page 308
CHANnel<m>:THReshold:FINDlevel on page 309
[RANGE] keys
The vertical [RANGE] keys set the vertical scale (vertical sensitivity) of the selected
waveform.
In FFT mode, the [RANGE] keys set the scale for the amplitude range (y-axis) in the
spectrum display.
In "Counter" mode, the [RANGE] keys set the measurement range.
Remote command:
CHANnel<m>:SCALe on page 304
CHANnel<m>:RANGe on page 304
FFT mode:
SPECtrum:FREQuency:MAGNitude:SCALe on page 357
Spectrum mode:
SPECtrum:FREQuency:SCALe on page 362
Counter mode:
COUNter<m>:SENSe:RANGe on page 415
[POS] keys
Move the selected signal up or down in the diagram. The position is a graphical setting
given in divisions, while the offset sets a voltage.
You can also drag the channel marker on the screen.
Remote command:
CHANnel<m>:POSition on page 306
Spectrum mode:
SPECtrum:FREQuency:POSition on page 362