Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
425User Manual 1326.1578.02 ─ 15
LOGGer:ZOOM:SCALe <Zoom Scale>
Sets the time scale of the zoomed waveform.
Parameters:
<Zoom Scale> AUTO | S1 | S2 | S4 | S5 | S10 | S20 | S40 | M1 | M2 | M4 | M5 |
M10 | M20 | M40 | H1 | H2 | H4 | H5 | H10 | H20 | D1 | D2 | D4
LOGGer:ZOOM:POSition <Time>
Sets or returns the position of the left edge of the zoomed area in relation to the left
side of the display.
Return values:
<Time> List of values:
<Year>,<Month>,<Day>,<Hour>,<Minute>,<Second>,
<TenthsOfSecond>
Year, month, day, hour, minute, second, tenth of a second, for
example 2016,10,29,16,10,22,2
Example: After 5 hours of logging, you want to see the first sample in the
zoom. Therefore, you enter the start time as zoom position. If
you scale the zoom, the start position remains.
15.10.5 Logger Statistics
The suffix <m> specifies the measurement that is logged.
LOGGer:RECording:STARt?........................................................................................... 425
LOGGer:RECording:TOTal?............................................................................................ 426
LOGGer:MEASurement<m>:RESult:CURRentsampl?....................................................... 426
LOGGer:MEASurement<m>:RESult:MAXimum:POSition?................................................. 426
LOGGer:MEASurement<m>:RESult:MAXimum:VALue?.....................................................426
LOGGer:MEASurement<m>:RESult:MINimum:POSition?.................................................. 427
LOGGer:MEASurement<m>:RESult:MINimum:VALue?......................................................427
LOGGer:MEASurement<m>:RESult:MEAN?.....................................................................427
LOGGer:MEASurement<m>:RESult:STDDev?..................................................................427
LOGGer:MEASurement<m>:ENABled?............................................................................428
LOGGer:MEASurement<m>:TYPE?................................................................................ 428
LOGGer:MEASurement<m>:SOURce?............................................................................ 428
LOGGer:RECording:STARt?
Returns the absolute start time of the current logging session.
Return values:
<StartTime> Year, month, day, hour, minute, second, tenth of a second, for
example 2015,10,29,16,10,22,2
Usage: Query only
Data Logging