Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
421User Manual 1326.1578.02 ─ 15
Setting parameters:
<Upper> Upper range limit
Example: Set the lower value to -1 and the upper value to +1 using both
commands:
LOGGer:MEASurement1:VERTical:LOWer -1
LOGGer:MEASurement1:VERTical:UPPer +1
Example: Set the lower value to -1 and the upper value to +1 using one
command:
LOGGer:MEASurement1:VERTical:LOWer -1, +1
//or
LOGGer:MEASurement1:VERTical:UPPer +1, -1
Firmware/Software: FW 1.60 and higher
LOGGer:MEASurement<m>:VERTical:DEViation <Deviation>
Sets the distance between the lower range and the mean value and between the upper
range and the mean value.
The command disables the vertical auto scaling.
See also: LOGGer:MEASurement<m>:VERTical:MEAN.
Suffix:
<m>
.
1..4
Logger channel
Parameters:
<Deviation> Absolute difference between mean value and upper/lower range
Range: 1E-15 to 100E+24
Increment: 1
*RST: 0.707
Firmware/Software: FW 1.60 and higher
LOGGer:MEASurement<m>:VERTical:MEAN <Mean>, [<Deviation>]
Sets the mean value of the scaling range. In additiion, you can set the deviation. The
query returns the mean value.
The command disables the vertical auto scaling.
Use this command if you want to define the range based on the mean value and the
standard deviation, see LOGGer:MEASurement<m>:RESult:MEAN? and LOGGer:
MEASurement<m>:RESult:STDDev? on page 427.
Suffix:
<m>
.
1..4
Logger channel
Parameters:
<Mean> Middle value between the upper and lower scaling value.
Data Logging