Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
420User Manual 1326.1578.02 ─ 15
15.10.2 Scaling
LOGGer:AUToset...........................................................................................................420
LOGGer:MEASurement<m>:VERTical:AUTO....................................................................420
LOGGer:MEASurement<m>:VERTical:UPPer...................................................................420
LOGGer:MEASurement<m>:VERTical:LOWer.................................................................. 420
LOGGer:MEASurement<m>:VERTical:DEViation.............................................................. 421
LOGGer:MEASurement<m>:VERTical:MEAN................................................................... 421
LOGGer:AUToset
Sets all logger channels to automatic scaling.
Usage:
Event
Firmware/Software: FW 1.60 and higher
LOGGer:MEASurement<m>:VERTical:AUTO [<State>]
Enables or disables the vertical auto scaling for the specified logger channel.
Using the command without parameter performs auto scaling once, without changing
the state.
Suffix:
<m>
.
1..4
Logger channel
Parameters:
<State> ON | OFF
*RST: ON
Example:
LOGGer:MEASurement1:VERTical:AUTO 0 //sets manual scaling for ch1
LOGGer:MEASurement1:VERTical:AUTO //autoscales logger channel 1
LOGGer:MEASurement1:VERTical:AUTO?
<-- 0 //state is unchanged (off)
Firmware/Software: FW 1.60 and higher
LOGGer:MEASurement<m>:VERTical:UPPer <Upper>, [<Lower>]
LOGGer:MEASurement<m>:VERTical:LOWer <Lower>, [<Upper>]
Set the upper and lower range limits for the specified logger channel. The values must
differ. You can set both values using only one of the commands (see examples). The
query returns the mandatory parameter.
The commands disable the vertical auto scaling.
Suffix:
<m>
.
1..4
Logger channel
Parameters:
<Lower> Lower range limit
Data Logging