Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
403User Manual 1326.1578.02 ─ 15
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Parameters:
<CoupVoltageMeas> DCLimit | ACLimit
DCLimit
Enables AC+DC measurement.
ACLimit
Enables AC measurement.
*RST: DCLimit
Example: Configure and perform AC+DC current measurement with range
100 A:
:METer:CONFigure:CURRent:AC 100
:METer:SENSe:FUNCtion 'CURR:AC'
:METer:CONFigure:VALue?
<-- "CURR:AC 100"
:METer:SENSe:CURRent:AC:COUPling DCL
:METer:READ?
<-- 0.035906488794
Configure and perform AC voltage measurement with range 10
V:
:METer:CONFigure:VAOLTage:AC 10
:METer:SENSe:FUNCtion 'VOLT:AC'
:METer:CONFigure:VALue?
<-- "VOLT:AC 10"
:METer:SENSe:CURRent:AC:COUPling ACL
:METer:READ?
<-- 0.030006488794
15.8.3 Relative Measurements
To perform relative measurements, set the relevant
METer<m>:SENSe:<function>:NULL:STATe command to ON. By default, the refer-
ence value is 0. Alternatively, you can use METer<m>:SENSe:RELative:STATe.
To change the reference value, use the
METer<m>:SENSe:<function>:NULL:VALue command.
METer<m>:SENSe:CURRent:AC:NULL:STATe................................................................. 404
METer<m>:SENSe:CURRent:DC:NULL:STATe................................................................. 404
METer<m>:SENSe:VOLTage:AC:NULL:STATe.................................................................. 404
METer<m>:SENSe:VOLTage:DC:NULL:STATe..................................................................404
METer<m>:SENSe:NULL:STATe......................................................................................404
METer<m>:SENSe:RELative:STATe.................................................................................404
METer<m>:SENSe:CURRent:AC:NULL:VALue................................................................. 404
METer<m>:SENSe:CURRent:DC:NULL:VALue................................................................. 404
Voltmeter (R&S