METer<m>:SENSe:RANGe <MeterRangeUI>
Sets the measurement range of the input channel that is measured by the meter.
Suffix:
<m>
.
1..4
Selects the voltmeter.
Parameters:
<MeterRangeUI> Range: The range depends on the selected measurement
type. For values, see the appropriate METer:CON-
Figure:<function> command.
15.8.2.2
METer<m>:CONFigure Commands
METer:CONFigure:<function> commands set all internal measurement parame-
ters for the specified measurement, and also the measurement range.
To set the range to minimum, maximum, or default value, use the following parameters:
METer<m>:CONFigure:<function> MIN
METer<m>:CONFigure:<function> MAX
METer<m>:CONFigure:<function> DEF
The METer<m>:CONFigure:<function> commands are only for configuration. To
activate a measurement, use METer<m>:SENSe:FUNCtion on page 405.
METer<m>:CONFigure:VALue? returns the active measurement and range.
To read the result, use METer<m>:READ? or METer<m>:FETCh?.
METer<m>:CONFigure:CURRent:AC............................................................................... 400
METer<m>:CONFigure:CURRent:DC...............................................................................400
METer<m>:CONFigure:VOLTage:AC................................................................................401
METer<m>:CONFigure:VOLTage:DC............................................................................... 401
METer<m>:CONFigure:CURRent:AC
METer<m>:CONFigure:CURRent:DC
Sets the internal parameters and configures the range for current measurements.
You need an external shunt resistor or I/U converter for current measurement.
Suffix:
<m>
.
1..4
Selects the voltmeter.
Voltmeter (R&S RTH1004)