Remote Control Commands
R&S
®
Scope Rider RTH
399User Manual 1326.1578.02 ─ 15
15.8.1 Activating the Voltmeter
METer<m>:SENSe:STATe <State>
Activates the voltmeter measurement.
Suffix:
<m>
.
1..4
Selects the voltmeter.
Parameters:
<State> ON | OFF
*RST: OFF
15.8.2 Measurement Configuration
The only parameter that can be set is the measurement range. All other settings are
automatically adjusted by the instrument.
To set a fixed measurement range, you can use several commands:
METer<m>:CONFigure:<function>
Configures the specified measurement including the measurement range.
See Chapter 15.8.2.2, "METer<m>:CONFigure Commands", on page 400.
METer<m>:SENSe:<function>:RANGe:UPPer
Sets the measurement range.
See Chapter 15.8.2.3, "METer<m>:SENSe:<function>:RANGe:UPPER Com-
mands", on page 401.
METer<m>:MEASure:<function>
Configures the specified measurement including the measurement range, starts
the measurement, and returns the result.
The configuration commands are described in the following chapters:
General Configuration...........................................................................................399
METer<m>:CONFigure Commands......................................................................400
METer<m>:SENSe:<function>:RANGe:UPPER Commands................................401
AC+DC Current and Voltage Measurements........................................................ 402
15.8.2.1 General Configuration
METer<m>:SENSe:SOURce <InputChannel>
Selects the input channel to be measured by the specified voltmeter.
Suffix:
<m>
.
1..4
Selects the voltmeter.
Parameters:
<InputChannel> C1 | C2 | C3 | C4
*RST: C1
Voltmeter (R&S RTH1004)