Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
398User Manual 1326.1578.02 ─ 15
Return values:
<Minimum> Range: -100E+24 to 100E+24
Increment: 1E-12
*RST: 0
Default unit: V
Usage: Query only
METer<m>:CALCulate:AVERage:MAXimum?
Returns the maximum value of the measurement series.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Return values:
<Maximum> Range: -100E+24 to 100E+24
Increment: 1E-12
*RST: 0
Default unit: V
Usage: Query only
METer<m>:CALCulate:AVERage:CLEar
Deletes all statistical values.
Statistics are also deleted if:
The measurement function changes (METer:SENSe:FUNCtion).
*RST
SYSTem:PRESet
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Usage: Event
15.8 Voltmeter (R&S RTH1004)
Activating the Voltmeter........................................................................................ 399
Measurement Configuration..................................................................................399
Relative Measurements........................................................................................ 403
Measurement Control............................................................................................405
Results.................................................................................................................. 407
Voltmeter (R&S
RTH1004)