Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
390User Manual 1326.1578.02 ─ 15
Parameters:
<Unit> C | F | K
METer:SENSe:TEMPerature:TRANsductor:RTD:TYPE <Unit>
Sets the adapter type for temperature measurements.
Parameters:
<Unit> PT100 | PT500
15.7.3 Relative Measurements
To perform relative measurements, set the relevant
METer:SENSe:<function>:NULL:STATe command to ON. By default, the reference
value is 0. Alternatively, you can use METer<m>:SENSe:RELative:STATe.
To change the reference value, use the METer:SENSe:<function>:NULL:VALue
command.
METer:SENSe:CAPacitance:NULL:STATe........................................................................ 390
METer:SENSe:CURRent:AC:NULL:STATe........................................................................390
METer:SENSe:CURRent:DC:NULL:STATe........................................................................390
METer:SENSe:RESistance:NULL:STATe.......................................................................... 390
METer:SENSe:TEMPerature:NULL:STATe........................................................................390
METer:SENSe:VOLTage:AC:NULL:STATe........................................................................ 390
METer:SENSe:VOLTage:DC:NULL:STATe........................................................................ 390
METer<m>:SENSe:NULL:STATe......................................................................................391
METer<m>:SENSe:RELative:STATe.................................................................................391
METer:SENSe:CAPacitance:NULL:VALue........................................................................ 391
METer:SENSe:CURRent:AC:NULL:VALue........................................................................391
METer:SENSe:CURRent:DC:NULL:VALue....................................................................... 391
METer:SENSe:VOLTage:AC:NULL:VALue........................................................................ 391
METer:SENSe:VOLTage:DC:NULL:VALue........................................................................ 391
METer:SENSe:RESistance:NULL:VALue..........................................................................392
METer:SENSe:TEMPerature:NULL:VALue........................................................................392
METer<m>:SENSe:NULL:VALU.......................................................................................392
METer:SENSe:CAPacitance:NULL:STATe <State>
METer:SENSe:CURRent:AC:NULL:STATe <State>
METer:SENSe:CURRent:DC:NULL:STATe <State>
METer:SENSe:RESistance:NULL:STATe <State>
METer:SENSe:TEMPerature:NULL:STATe <State>
METer:SENSe:VOLTage:AC:NULL:STATe <State>
METer:SENSe:VOLTage:DC:NULL:STATe <State>
Enables or disables the relative measurement. The reference value is defined using
the appropriate METer:SENSe:<function>:NULL:VALue command.
Digital Multimeter (R&S
RTH1002)