386User Manual 1326.1578.02 ─ 15
Parameters:
<Range> <numeric value> | 'AUTO' | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 1 V|10 V|100 V|
1000 V.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Range: 1 to 1000
*RST: 1
Default unit: V
Usage: Setting only
15.7.2.2 METer:SENSe:<function>:RANGe:UPPER Commands
METer:SENSe:<function>:RANGe:UPPER commands set the measurement range
and turn off autoranging. You can use these commands in addition to
METer:CONFigure commands if you want to change only the range.
METer:SENSe:CAPacitance:RANGe:UPPer..................................................................... 386
METer:SENSe:CURRent:AC:RANGe:UPPer.....................................................................386
METer:SENSe:CURRent:DC:RANGe:UPPer.....................................................................386
METer:SENSe:FREQuency:VOLTage:RANGe:UPPer........................................................ 387
METer:SENSe:RESistance:RANGe:UPPer....................................................................... 387
METer:SENSe:VOLTage:AC:RANGe:UPPer..................................................................... 387
METer:SENSe:VOLTage:DC:RANGe:UPPer..................................................................... 387
METer:SENSe:CAPacitance:RANGe:UPPer <Range>
Sets a fixed range for capacitance measurements and turns off auto ranging.
Parameters:
<Range> <numeric value> | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 10 nF|100 nF|
1 μF|10 μF|100 μF|1 mF|10 mF.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Range: 10 nF to 10 mF
*RST: 10 nF
METer:SENSe:CURRent:AC:RANGe:UPPer <Range>
METer:SENSe:CURRent:DC:RANGe:UPPer <Range>
Sets a fixed current range and turns off auto ranging.
Digital Multimeter (R&S
RTH1002)