Parameters:
<OperationMode> YT | XY | MASK | ROLL | METer | FFT | LOGGer | COUNter |
PROTocol | HARMonic | SPECtrum
*RST: YT (scope mode)
15.7.2 Measurement Configuration
The instrument sets most configuration settings automatically. For most measurement
types, the measurement range is the only parameter that can be set. For temperature
measurements, specific settings are required.
The instrument can adjust the measurement range if autoranging is configured for a
measurement. Otherwise, you can set a fixed measurement range.
To set a fixed measurement range, you can use several commands:
METer:CONFigure:<function>
Configures the specified measurement including the measurement range.
See Chapter 15.7.2.1, "METer:CONFigure Commands", on page 383.
METer:SENSe:<function>:RANGe:UPPer
Sets the measurement range and turns off autoranging.
See Chapter 15.7.2.2, "METer:SENSe:<function>:RANGe:UPPER Commands",
on page 386.
METer:MEASure:<function>
Configures the specified measurement including the measurement range, starts
the measurement, and returns the result.
To enable autoranging, you can use the following commands:
METer:SENSe:<function>:RANGe:AUTO
Enables or disables the autoranging. See Chapter 15.7.2.3, "METer:SENSe:<func-
tion>:RANGe:AUTO Commands", on page 388.
METer:CONFigure:<function> 'AUTO'
See Chapter 15.7.2.1, "METer:CONFigure Commands", on page 383.
The configuration commands are described in the following chapters:
METer:CONFigure Commands............................................................................. 383
METer:SENSe:<function>:RANGe:UPPER Commands.......................................386
METer:SENSe:<function>:RANGe:AUTO Commands......................................... 388
AC+DC Current and Voltage Measurements........................................................ 389
Temperature Measurements................................................................................. 389
15.7.2.1 METer:CONFigure Commands
METer:CONFigure:<function> commands set all internal measurement parame-
ters for the specified measurement. For most measurements, it also sets the measure-
ment range.
To set the range to minimum, maximum, or default value, use the following parameters:
METer:CONFigure:<function> MIN
Digital Multimeter (R&S
RTH1002)