Return values:
<Phase> Default unit: degrees
Usage: Query only
HARMonic:RESult<m>:HARMonics<n>:RELMagnitude:MAXimum?
HARMonic:RESult<m>:HARMonics<n>:RELMagnitude:MINimum?
HARMonic:RESult<m>:HARMonics<n>:RELMagnitude[:CURRent]?
Returns the currently measured, minimum or maximum power level of the selected
harmonic, relative to the power level of the fundamental frequency.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
<n> 1..64
Order of harmonic (possibly restricted, see HARMonic:
RESult<m>:NOHarmonics?)
Return values:
<RelMagnitude> Default unit: %
Usage: Query only
HARMonic:RESult<m>:HARMonics<n>[:FREQuency]?
Queries the determined frequency value for the specified harmonic.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
<n> 1..64
Order of harmonic (possibly restricted, see HARMonic:
RESult<m>:NOHarmonics?)
Return values:
<Frequency> Default unit: Hz
Usage: Query only
HARMonic:RESult<m>:LIMViolation?
Queries the result of the limit check for all harmonics. If a single harmonic violates its
defined limit, the limit check for the entire signal fails.
Note: Only if a limit value is defined, a limit check is performed for the harmonic (see
HARMonic:RESult<m>:HARMonics<n>:LIMCheck? on page 377).
Tip: To query the result of the limit check for an individual harmonic, use HARMonic:
RESult<m>:HARMonics<n>:LIMViolation? on page 379.