Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
378User Manual 1326.1578.02 ─ 15
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
<n> 1..64
Order of harmonic (possibly restricted, see HARMonic:
RESult<m>:NOHarmonics?)
Return values:
<State> ON | OFF
ON
A limit is specified and a limit check will be performed.
OFF
No limit is specified, so no limit check will be performed for the
harmonic.
Example:
HARM:LIM:TYPE EN50160
HARM:RES1:HARM2:LIMC?
//Result: 1
HARM:RES1:HARM2:LIM?
//Result: 2%
HARM:RES1:HARM2:LIMV?
//Result: 0
Usage: Query only
HARMonic:RESult<m>:HARMonics<n>:LIMit?
Queries the limit defined for the specified harmonic.
Note: Only if a limit value is defined, a limit check is performed for the harmonic (see
HARMonic:RESult<m>:HARMonics<n>:LIMCheck? on page 377).
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
<n> 1..64
Order of harmonic (possibly restricted, see HARMonic:
RESult<m>:NOHarmonics?)
Return values:
<Limit> percentage
Example:
HARM:LIM:TYPE EN50160
HARM:RES1:HARM2:LIMC?
//Result: 1
HARM:RES1:HARM2:LIM?
//Result: 2%
HARM:RES1:HARM2:LIMV?
//Result: 0
Usage: Query only
Spectrum Analysis