Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
372User Manual 1326.1578.02 ─ 15
Harmonics that were previously eliminated by the HARMonic:LIMits:TYPE com-
mand cannot be displayed.
Parameters:
<Selection> ALL | EVEN | ODD | ODD3 | ODN3 | USER
ODD3
All harmonics with an odd-numbered order that is a multiple of 3
ODN3
All harmonics with an odd-numbered order that cannot be divi-
ded by 3
USER
Selects a user-defined number of harmonics. Define the number
of harmonics using the HARMonic:DISPlay:USER command.
*RST: ALL
HARMonic:DISPlay:USER
<MaxHarmonics>
Determines the maximum number of harmonics to be displayed in the bargraph. This
command requires a preceding HARM:DISP:TYPE USER command.
Note that this setting only affects the graphical result display; it has no effect on the
selection of harmonics for which results are provided during a file export (manually or
via remote command). For that purpose, use HARMonic:LIMits:TYPE on page 374.
Harmonics that are eliminated by the HARMonic:LIMits:TYPE command cannot be
displayed.
Parameters:
<MaxHarmonics> Range: 1 to 64
Increment: 1
*RST: 10
Example:
HARM:DISP:TYPE USER
HARM:DISP:USER 25
Displays the first 25 harmonics.
HARMonic:FUNDamental:TYPE <Fundamental>
Defines the basis of the harmonics measurement. Harmonics are determined as multi-
ples of this frequency.
Parameters:
<Fundamental> F50 | F60 | F400 | USER
F50
50 Hz
F60
60 Hz
F400
400 Hz
Spectrum Analysis