Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
366User Manual 1326.1578.02 ─ 15
Return values:
<LevelDelta> Range: -520 to 520
Increment: 1
*RST: 0
Default unit: dB
Usage: Query only
SPECtrum:CURSor<m>:SCReen
Sets the cursors to a default position on the screen.
Suffix:
<m>
.
1..2
Usage: Event
15.6.2.3
Markers in Spectrum Mode
Special marker commands are available for the Spectrum mode.
SPECtrum:MARKer[:STATe]............................................................................................366
SPECtrum:MARKer:COUNt............................................................................................ 366
SPECtrum:MARKer:SOURce.......................................................................................... 367
SPECtrum:MARKer:SETup:DISTance.............................................................................. 367
SPECtrum:MARKer:SETup:EXCursion.............................................................................367
SPECtrum:MARKer:SETup:MLEVel................................................................................. 367
SPECtrum:MARKer:RCOunt?......................................................................................... 368
SPECtrum:MARKer:RESult<m>:FREQuency[:VALue]?......................................................368
SPECtrum:MARKer:RESult<m>:FREQuency:DELTa?........................................................368
SPECtrum:MARKer:RESult<m>:LEVel[:VALue]?............................................................... 369
SPECtrum:MARKer:RESult<m>:LEVel:DELTa?.................................................................369
SPECtrum:MARKer:RMARker:FREQuency?.....................................................................369
SPECtrum:MARKer:RMARker:VALue?.............................................................................370
SPECtrum:MARKer[:STATe] <State>
If enabled, a peak search is performed on the current spectrum results.
Parameters:
<State> ON | OFF
*RST: OFF
SPECtrum:MARKer:COUNt <NumberOfMarkers>
Defines the number of markers used to indicate peaks in the spectrum results. With 3
active markers, the 3 peak values in the spectrum are indicated.
Spectrum Analysis