Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
362User Manual 1326.1578.02 ─ 15
Parameters:
<RefLevel> Range: -160 to 160
Increment: 1
*RST: 0
Default unit: dB
SPECtrum:FREQuency:POSition <Frequency>
Configures the position of the spectrum within the vertical diagram grid. Changing this
value has the same effect as using the [POS] keys on the instrument.
Note that this command contains the keyword FREQuency for compatibility reasons. In
effect, it changes the level position in the spectrum diagram.
Parameters:
<Frequency> Defines the number of divisions in the vertical grid that the spec-
trum is moved up (positive value) or down (negative value).
Range: -4 to 4
Increment: 0.5
*RST: 2
Default unit: none
SPECtrum:FREQuency:SCALe <VerticalScale>
Configures the scale of the y-axis of the spectrum, which is indicated in the channel
settings beneath the spectrum diagram. Changing this value has the same effect as
using the [RANGE] keys on the instrument.
Note that this command contains the keyword FREQuency for compatibility reasons. In
effect, it changes the level scaling in the spectrum diagram.
Parameters:
<VerticalScale> Defines the level range displayed in one division of the vertical
grid.
Range: 0.5 to 40
Increment: 0.1
*RST: 10
Default unit: dB
Usage: Asynchronous command
SPECtrum:WAVeform:AVERage[:ENABle] <State>
Displays the averaged spectrum trace. The number of traces to average is defined
using SPECtrum:FREQuency:AVERage:COUNt on page 360.
Parameters:
<State> ON | OFF
*RST: OFF
Spectrum Analysis