Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
353User Manual 1326.1578.02 ─ 15
Suffix:
<m>
.
1|2|5 (RTH1002) , 1..5 (RTH1004)
Parameters:
<WidthX> Range: 0 to 10
Increment: 0.01
*RST: 0.05
Default unit: div
MASK:CHANnel<m>:PROPerties:YWIDth <WidthY>
Changes the width of the selected mask in vertical direction.
Suffix:
<m>
.
1|2|5 (RTH1002) , 1..5 (RTH1004)
Parameters:
<WidthY> Range: 0 to 8
Increment: 0.01
*RST: 0.1
Default unit: div
MASK:CHANnel<m>:CREatemask
Creates a mask from the envelope waveform of the selected waveform with the
defined width in x and y direction.
Suffix:
<m>
.
1|2|5 (RTH1002) , 1..5 (RTH1004)
Usage: Event
15.5.2 Mask Test
MASK:ONViolation[:SELection]....................................................................................... 353
MASK:RST....................................................................................................................354
MASK[:TESTstate]?........................................................................................................354
MASK:CHANnel<m>:RESult:FAIL:PERCentage?.............................................................. 354
MASK:CHANnel<m>:RESult:FAIL[:COUNt]?.....................................................................354
MASK:CHANnel<m>:RESult:PASS:PERCentage?............................................................ 355
MASK:CHANnel<m>:RESult:PASS[:COUNt]?................................................................... 355
MASK:CHANnel<m>:RESult:TOTL[:COUNt]?................................................................... 355
MASK:ELAPsedtime:TOTal?........................................................................................... 355
MASK:ELAPsedtime[:SECS]?......................................................................................... 356
MASK:ONViolation[:SELection] <SelectedActions>
Defines the action to be executed if a violation occurs.
Mask Testing